Detail Radon Meter DURRIDGE RAD7
Modes of Operation
SNIFF Rapid response and rapid recovery radon measurement
THORON Radon and thoron measured simultaneously and independently
NORMAL High sensitivity
AUTO Automatic switch from SNIFF to NORMAL after three hours run
GRAB Analysis of grab samples
WAT Automatic analysis of water samples with RAD H2O accessory
Measurements
Radon in air with Sniff protocol for quick, spot reading
Thoron protocol for searching for radon entry points
Radon in air 1-day, 2-day or weeks protocol for long term measurement
Radon in water batch samples with RAD H2O and Big Bottle RAD H2O
Continuous radon in water with RAD AQUA and Radon-in-Water Probe
Radon in soil gas with Soil Gas Probe and Active DRYSTIK
Radon emission from soil and hard surfaces with surface emission chamber
Bulk radon emission from bulk materials and objects
Data Storage
1, 000 records, each with 23 fields of data
Log of printer output also stored
Sample Pumping
Built-in pump draws sample from chosen sampling point
Flow rate typically 800ml/ min
Print Output
Short, medium or long format data printed after each cycle
Run summary printed at end of run, including averages and spectrum
PC Connectivity
RS232 serial port, full remote control implemented in CAPTURE Software.
Optional serial to Bluetooth adaptor for wireless PC connectivity.
Audio Output
GEIGER Tone beeps for radon and thoron counts
CHIME Chime only at the end of each cycle, otherwise silent
OFF No sound
Tamper Resistance
TEST LOCK command locks keypad to secure against tampering
Technical Specifications:
Principle of Operation
Electrostatic collection of alpha-emitters with spectral analysis
Passivated Ion-implanted Planar Silicon detector
SNIFF mode counts polonium-218 decays
NORMAL mode counts both polonium 218 and polonium 214 decays
Built-In Air Pump
Nominal 1 liter/ minute flow rate
Inlet and outlet Luer connectors
Connectivity
RS-232 port up to 19, 200 baud rate
USB adaptor is included with every RAD7
Measurement Accuracy
+ / -5% absolute accuracy, 0% - 100% RH
Nominal Sensitivity
SNIFF mode, 0.25 cpm/ ( pCi/ L) , 0.0067 cpm/ ( Bq/ m3)
NORMAL mode, 0.5 cpm/ ( pCi/ L) , 0.013 cpm/ ( Bq/ m3)
Radon Concentration Range
0.01 - 20, 000 pCi/ L ( 0.4 - 750, 000 Bq/ m3)
Intrinsic Background
0.005 pCi/ L ( 0.2 Bq/ m3) or less, for the life of the instrument
Recovery Time
Residual activity in Sniff mode drops by factor of 1, 000 in 30 minutes
Operating Ranges
Temperature: 32° - 113° F ( 0° - 45° C)
Humidity: 0% - 100% , non-condensing
Cycle Range
User controllable number of cycles, from 1 to 99 to unlimited, per run
User controllable cycle time, from 2 minutes to 24 hours
CAPTURE Software
Compatible with Microsoft Windows XP and 7, and Mac OS X
Automatic RAD7 location, connection and data download
Graphs radon, thoron, temperature and humidity over time
Automatic humidity correction
Statistical analysis tools track concentration averages and uncertainties
Chart Recorder mode provides real-time RAD7 status monitoring
Control RAD7 operations from computer via direct or remote connection
Automatic calculation and display of radon in water for RAD AQUA
Automatic combination of multiple RAD7 data
Physical Specifications:
Dimensions
11.5" x 8.5" x 11" ( 29.5 cm x 21.5 cm x 27.9 cm)
Weight
9.6 pounds ( 4.35 kg)
LCD Display Output
2 line x 16 character alpha-numeric display
Case Material
High density polyethylene
Infrared Printer
Chamjin NewHandy 700 Wireless Infrared Printer included
Power Supply
11-15V DC ( 12V nominal) @ 1.25A, center pin positive, or internal rechargeable batteries ( 5Ah)
Optional AC input for non-EU customers
Battery Longevity
24 hours in SNIFF mode; 72 hours in Monitor mode